Table 1 Intrasession variability characteristics and compliance rates for the FICO and FICH4 of Masterscreen and MeAir

From: The Reliability of a Novel Structured Testing Device for Single-breath Lung Carbon Monoxide Uptake: A Randomized Comparison Crossover Study

Gas

Variables

MasterScreen

MeAir

FICO

Measured value

0.295(0.291,0.299)

0.298(0.296,0.300)

Min-Max

0.262–0.310

0.287–0.305

CV

2.6%

1.2%

Measurement error

-12.667%∼1.667%

-4.333%∼1.667%

The proportion of acceptable

57/160(35.63%)

101/160(63.13%)

FICH4

Measured value

0.301(0.297,0.305)

0.299(0.295,0.301)

Min-Max

0.266–0.314

0.287–0.304

CV

3.0%

1.2%

Measurement error

-11.333%∼4.667%

-4.333%∼1.333%

The proportion of acceptable

69/160(43.13%)

107/160(66.88%)

  1. The target gas concentration is 0.3%. As the data distribution of inhaled gas (both the two devices) do not conform to the normal distribution, so data are presented as median (P25,P75) and minimum to maximum. FICO (%): fractional concentration of Inspired carbon monoxide; FICH4 (%): fractional concentration of Inspired methane; CV: coefficient of variation; The measurement error is calculated by the difference between the minimum gas concentration and maximum gas concentration to the target gas concentration as a percentage of the target gas concentration. The proportion of acceptable is the ratio where the measurement error is within 1%