The Reliability of a Novel Structured Testing Device for Single-breath Lung Carbon Monoxide Uptake: A Randomized Comparison Crossover Study

Abstract

Background

With the development and market launch of several new domestic lung diffusing capacity testing instruments, the clinical reliability of the MeHow MeAir 9000 spirometer, featuring a novel turning valve structure, needs to be validated.

Objective

To evaluate the clinical reliability of lung diffusing capacity measurements using the MeHow MeAir 9000 spirometer.

Methods

This study included 166 participants: 30 healthy individuals, 68 with interstitial lung disease (ILD), and 68 with chronic obstructive pulmonary disease (COPD). Using a crossover design, participants underwent lung diffusing capacity tests with both the MeHow MeAir 9000 and Jaeger MasterScreen Diffusion spirometers, following the 2017 ERS/ATS standards. The primary indicator was the diffusing capacity for carbon monoxide (DLCO), with the diffusion capacity of carbon monoxide as a percentage of the predicted value (DLCO %pred) as the main categorical indicator. Secondary indicators included DLCO to alveolar volume ratio (DLCO / VA), alveolar volume (VA), inspired volume (VI), breath-hold time (tBH), fractional concentration of inhaled carbon monoxide (FICO), fractional concentration of inhaled methane (FICH4), fractional concentration of carbon monoxide in the alveolar space (FACO), and fractional concentration of methane in the alveolar space (FACH4). Consistency analysis was performed on the measurements and the classification of lung diffusing capacity impairment severity from both instruments. Additionally, scatter plots and coefficient of variation (CV%) for inhaled carbon monoxide (CO) and methane (CH4) concentrations were analyzed, along with simulator (Hans Rudolph, Kansas City, MO) test results.

Results

DLCO ‘s Bland-Altman plot showed 95.0% of data points within the 95% CI, with a CCC of 0.988. In the DLCO %pred classification, the ICCs for the ILD group and the COPD group were 0.940 and 0.975, respectively, while the ICC for the healthy subject group was 0.931. These results indicate good consistency for the primary outcome measures. Secondary indicators had varying CCCs, indicating fair to poor consistency (P < 0.05). Scatter plots and CV% for inhaled CO and CH4 concentrations suggested better stability of MeAir over MasterScreen. Simulator test results showed MeAir had a CO error range of -3.80% to -1.00% and CH4 error range of -0.50–1.00%, while MasterScreen had a CO error range of -2.53–0.00% and CH4 error range of -1.83% to -0.63%, indicating superior CO detection by MasterScreen and better CH4 detection by MeAir.

Conclusion

The MeHow MeAir 9000 spirometer provides high accuracy measurements of lung diffusing capacity and reliable assessment of the severity of diffusing capacity impairment, making it suitable for clinical use.

Clinical trial number

Not applicable.

Peer Review reports

Introduction

Pulmonary diffusion capacity refers to the process where gases such as oxygen diffuse through the alveolar-capillary membrane from the alveoli to the capillaries, eventually entering the bloodstream and binding with hemoglobin in red blood cells. Clinically, carbon monoxide(CO) is commonly used as a test gas to assess this function [1], and in diffusion function tests, the tracer gases commonly used are helium (He) and methane (CH4) [2]. The key indicator of diffusion function testing is the Diffusing Capacity of the Lung for Carbon Monoxide (DLCO). The factors affecting it include the concentration of the test gas as well as the inspired volume (VI), breath-hold time (tBH), environmental factors (including ambient temperature (T), barometric pressure (PB), and the partial pressure of water vapor (PH2O)), and alveolar volume (VA) [2, 3].

Diffusion function testing of the lungs is widely used in the medical field to assess the condition of lung diseases [4,5,6]. As the application scope of this test continues to expand, it has been used for diagnosing and monitoring early chronic obstructive pulmonary disease (COPD) [7], distinguishing COPD from asthma [8], and evaluating the therapeutic effects and predicting the prognosis of interstitial lung disease (ILD) [9]. Therefore, the importance of lung diffusion function testing is increasingly highlighted.

Currently, although the market relies heavily on imported lung diffusion function testing equipment, especially the Jaeger MasterScreen Diffusion device (Vyaire Medical, Germany) (hereafter referred to as MasterScreen), which is widely used for its high reliability and authority, domestic lung function equipment is also continuously improving. Now, there are domestic lung function testing systems capable of performing lung diffusion function tests. Unlike the traditional imported MasterScreen, the new domestic MeHow MeAir 9000 device (MeHow Medical Technology Co., China) (hereafter referred to as MeAir) adopts an innovative valve-turning diffusion function module and is equipped with a software system that assists with an exhalation platform and indicator lines. The system can display the auxiliary line of the expiratory platform and automatically control the steering valve when the expiratory flow velocity drops to 0.025 L/s. When the residual gas volume is reached, the air valve is closed, and the diffusion test gas valve is opened. This system can not only standardize the standard of cooperative inspection between the operator and the subject, but also ensure the stability of the inhaled gas concentration, to ensure the accuracy of the measurement of pulmonary dispersion function. The purpose of this study is to comprehensively evaluate the applicability of the MeAir lung diffusion function system in different populations. We plan to perform lung diffusion function tests with the MeAir lung function device and the German MasterScreen lung diffusion function device in populations with restrictive ventilatory dysfunction, obstructive ventilatory dysfunction, respiratory diseases, and healthy individuals. The results will be analyzed and compared to assess the effectiveness of this new lung function instrument’s diffusion function system in clinical use.

Methods

According to the established inclusion and exclusion criteria, a single-center randomized crossover trial was conducted among pulmonary function test subjects at the First Affiliated Hospital of Guangzhou Medical University from June 15, 2023, to September 1, 2023.

Participants

The plan is to recruit 166 participants, divided into a COPD group with 68 cases, an ILD group with 68 cases, and a healthy control group with 30 cases who have no pulmonary diseases. There are no restrictions on the severity or the state of respiratory diseases of the participants. According to the 2017 European Respiratory Society/American Thoracic Society (ERS/ATS) technical standards for lung diffusion function testing, participants who meet the indications for pulmonary function testing and are excluded from the contraindications for pulmonary function testing will be enrolled [10]. According to the randomized crossover procedure, lung diffusion function tests will be completed using both the MasterScreen device and the MeAir device.

Randomization and single-blind

We used simple randomization to assign study participants to begin measurements with either the MasterScreen or MeAir device. A computer-generated list of random numbers was created by an independent person not involved in the study, using the randomization tool in SPSS version 26.0. The list contained a total of 166 numbers, which were either 1 or 2. The number 1 indicated that the participant started with the MasterScreen device, while the number 2 indicated that the participant started with the MeAir device. In this study, a single-blind method was also used for the examination of participants, meaning that the participants were unaware of their randomization numbers under which the tests were conducted. Additionally, in the study design, participants were eventually able to learn the results of their own pulmonary function test from either lung function device.

Quality control

Before the start of the study, both lung function devices underwent technical checks and rigorous quality control by the technicians or service providers. To ensure the acquisition of high-quality measurement results, we calibrated both devices according to the 2017 ERS/ATS technical standards for lung diffusion function testing before each day’s measurements [10]. Before starting the examination each day, we first calibrated the environmental factors such as PB, T, and PH2O. Subsequently, both lung function devices were manually calibrated using three-flow measurements and a standard 3.0 L syringe. In addition to volume calibration, the gas concentrations of the test gases, carbon monoxide and methane, were also calibrated.

All pulmonary function tests were completed by 9 senior technicians specialized in pulmonary function, with each participant’s two tests conducted by the same professional. The testing process was carried out according to the 2017 ERS/ATS technical standards for lung diffusion function testing [10].

Measurement protocol

All measurements were conducted on both MasterScreen and MeAir devices. During the assessment of lung ventilation and diffusion functions, participants were asked to maintain a seated position to avoid any effects on diffusion capacity measurements due to changes in cardiac output [11, 12]. In accordance with the 2017 ERS/ATS technical standards for lung diffusion function testing, the quality control criteria for the lung diffusion function test are as follows [10]:

Acceptability

  • The VI must be ≥ 85% of the vital capacity (VC).

  • The VA must be within 200 ml or 5% of the VA from the previous acceptable diffusion maneuver (whichever is larger).

  • The inhalation of the test gas to achieve 85% must be completed within 4.0 s.

  • The tBH should be 10 ± 2 s without leaks, Muller maneuvers (increasing negative thoracic pressure by forcefully inhaling with the glottis closed), and Valsalva maneuvers (increasing positive thoracic pressure by forcefully exhaling with the glottis closed). The oral pressure change during the breath hold should not exceed ± 3 kPa.

  • The exhalation time should be less than 4.0 s to exclude anatomical dead space and allow for appropriate alveolar gas sampling.

Repeatability

  • At least two acceptable DLCO measurements should be within 2 mL·min− 1·mmHg− 1 of each other.

  • No more than five measurements should be taken.

  • The inhalation of the test gas to achieve 85% must be completed within 4.0 s.

  • There should be at least a 4-minute interval between two measurements on the same spirometer.

Study end-points

To analyze the consistency of the diffusion function systems of the two lung function devices, we need to collect and analyze factors that affect the outcomes. According to the formulas for calculating DLCO and VA (Eqs. 1 and 2), we can understand that the main influencing factors include: volume of equipment dead (Vdequip), volume of anatomical dead (Vdanat), fractional concentration of inhaled carbon monoxide (FICO), fractional concentration of inhaled methane (FICH4), fractional concentration of carbon monoxide in the alveolar space (FACO), fractional concentration of methane in the alveolar space (FACH4), and environmental factors such as T and PB.

It is worth noting that there are some unavoidable human factors during the diffusion function check that may lead to uncontrollable differences in tBH between MeAir and MasterScreen, and we have indeed found such a situation in our statistics (Table 1). Although tBH is an influencing factor in the DLCO calculation formula, our previous research has found that after a breath-hold time of 8 s, the DLCO results show no significant relationship with tBH [13, 14]. Moreover, by ensuring that all participants have a stable breath-hold time of over 8 s, tBH does not affect the comparison of the main observation index, DLCO, between the two devices.

$$\:\begin{array}{c}{D}_{\text{L}\text{C}\text{O}}=\frac{{V}_{\text{A}\text{S}\text{T}\text{P}\text{D}}}{{t}_{\text{B}\text{H}}\cdot\:\left({P}_{\text{B}}-47\right)}\cdot\:\text{ln}\left(\frac{{F}_{\text{I}\text{C}\text{O}}}{{F}_{\text{A}\text{C}\text{O}}}\cdot\:\frac{{F}_{\text{A}\text{T}\text{r}}}{{F}_{\text{I}\text{T}\text{r}}}\right)\cdot\:60,000\end{array}$$
(1)
$\eqalign{{V_{{\rm{ABTPS}}}} = & \left({{V_{{\rm{IATPD}}}} - {V_{{\rm{Dequip}}}} - {V_{{\rm{Danat}}}}} \right) \cr & \cdot {{{F_{{\rm{ITr}}}}} \over {{F_{{\rm{ATr}}}}}} \cdot {{{P_{\rm{B}}}} \over {\left({{P_{\rm{B}}} - 47} \right)}} \cdot {{310} \over {\left({273 + T} \right)}} \cr}$
(2)

To integrate the clinical application of lung diffusion function tests, we referred to international technical guidelines for lung diffusion function and graded the diffusion dysfunction severity of the DLCO %pred measured in the two disease groups into four levels: normal (DLCO %pred ≥ 80%), mild (80% > DLCO %pred ≥ 60%), moderate (60% > DLCO %pred ≥ 40%), and severe (40% > DLCO %pred) [10].

Ultimately, we identified the main quantitative observation indicators for comparing the accuracy of the diffusion checks of the two lung function devices as follows: DLCO; secondary quantitative observation indicators include: DLCO to alveolar volume ratio (DLCO /VA), VA, FICO, FICH4, FACO, FACH4, tBH, and VI.

The main qualitative observation indicator for assessing the diffusion dysfunction of pulmonary disease is the grade of diffusion dysfunction based on the percentage of the measured carbon monoxide diffusing capacity to the predicted value (DLCO %pred).

Statistical analysis and sample size calculation

We utilized MedCalc and SPSS 26.0 statistical software for data analysis. The normality of the data was tested using the Kolmogorov-Smirnov test. If the data conform to a normal distribution, the quantitative data are represented as (x ± s); if it does not conform to a normal distribution, they are represented as M (P25, P75).

For quantitative data such as test results, we used the concordance correlation coefficient (CCC) to analyze and compare the results of the Primary and Secondary end-points. Bland-Altman plots were also analyzed the Primary end-point with bias estimates and 95% limits of agreement.

The interpretation of correlation from the CCC is as follows: a value less than 0.90 indicates poor strength of agreement, 0.90–0.95 indicates moderate strength of agreement, 0.95–0.99 indicates high strength of agreement, and greater than 0.99 indicates almost perfect agreement. Additionally, if the 95% confidence interval of the result includes 0, it indicates that the result is not statistically significant [15].

For categorical data such as the severity of disease diffusion, we used the intraclass correlation coefficient (ICC) (model: two-way mixed; type: absolute agreement) for analysis and comparison. Values less than 0.5 indicate poor reliability, values between 0.5 and 0.75 indicate moderate reliability, values between 0.75 and 0.9 indicate good reliability, and values greater than 0.90 indicate excellent reliability, with p < 0.05 considered statistically significant [16]. Additionally, we compared the recognition ability of the two lung function devices for the concentration of inhaled gases through the central tendency of the scatter plots and by calculating the coefficient of variation (CV) using the root mean square (RMS) method.

With the CCC for Power set at 0.95 and the Lower Boundary at 0.9, and the statistical power reaching 0.99, we need to include at least 124 participants for the analysis of the observation indicators (calculated using PASS 2021). Furthermore, considering a dropout rate of 20%, our goal is to recruit at least 150 participants in total.

Results

This study included a total of 136 patients and 30 normal individuals (Fig. 1). Among them, 68 were patients with ILD and 68 were patients with COPD. Among these patients, one ILD patient was excluded due to low lung capacity (< 1 L) which resulted in an inability to tolerate the breath-holding time, one COPD patient was excluded due to poor cooperation as they could not provide valid data on the Jaeger device, and one COPD patient withdrew midway and did not complete the test. Among the 30 normal individuals, 3 participants were unable to complete the examination due to poor cooperation.Therefore, a total of 160 participants’ data were included in the analysis, and no adverse events occurred throughout the study process. The characteristics of the study population are shown in Table 2.

Fig. 1
figure 1

MeAir flow chart

Table 1 Intrasession variability characteristics and compliance rates for the FICO and FICH4 of Masterscreen and MeAir
Table 2 Baseline demographic and clinical characteristics by sequence and by total

In this clinical study, there were 67 participants in the interstitial lung disease group with respiratory diseases, including 36 male participants (53.73%) and 31 female participants (46.27%), with an average age of 54.6 ± 12.9 years. The COPD group had 66 participants, including 59 male participants (89.39%) and 7 female participants (10.61%), with an average age of 63.3 ± 7.9 years. The health control group consisted of 27 participants, including 17 males (62.96%) and 10 females (37.04%), with an average age of 29.6 ± 5.7 years.

The Bland-Altman plot (Fig. 2) was used to perform a consistency statistical analysis of the main diffusion function observation indicator, DLCO test results. The results showed that 95.0% (152/160) of the observations were within the 95% confidence interval (CI). This indicates a high level of agreement between the measurements obtained from the two lung function devices, as most data points fall within the expected range of consistency.

Fig. 2
figure 2

Comparison of raw data distribution for Bland-Altman scatter plot of DLCO between MasterScreen and MeAir

The main observation indicator DLCO and the distribution of secondary observation indicators on the MeAir 9000 lung function device and the MasterScreen lung function device, as well as their consistency correlation coefficient statistical analysis, are detailed in Table 1.

Data conform to a normal distribution, the quantitative data are represented as (x ± s); if it does not conform to a normal distribution, it is represented by M (P25, P75). Absolute value of difference is calculated by the difference between the data in accordance with normal distribution unless otherwise stated. DLCO: diffusing capacity of the lung for carbon monoxide; DLCO/VA: transfer coefficient of the lung for CO; VA: alveolar volume; VI: inspired volume; tBH: Breath-hold time; VA/tBH: ratios of alveolar volume to breath-hold time; FICO (%): fractional concentration of Inspired carbon monoxide; FICH4 (%): fractional concentration of Inspired methane; FACO (%): fractional concentration of carbon monoxide in the alveolar space; FACH4 (%): fractional concentration of methane in the alveolar space; \(\:\frac{{F}_{\text{I}\text{C}\text{O}}}{{F}_{\text{A}\text{C}\text{O}}}\): ratio of inspired to expired carbon monoxide concentration; \(\:\frac{{F}_{\text{A}\text{C}\text{H}4}}{{F}_{\text{I}\text{C}\text{H}4}}\): ratio of expired to inspired methane concentration; \(\:\frac{{F}_{\text{I}\text{C}\text{O}}}{{F}_{\text{A}\text{C}\text{O}}}\cdot\frac{{F}_{\text{A}\text{C}\text{H}4}}{{F}_{\text{I}\text{C}\text{H}4}}\): Product of \(\:\frac{{F}_{\text{I}\text{C}\text{O}}}{{F}_{\text{A}\text{C}\text{O}}}\) and \(\:\frac{{F}_{\text{A}\text{C}\text{H}4}}{{F}_{\text{I}\text{C}\text{H}4}}\); \(\:\text{L}\text{n}\left(\frac{{F}_{\text{I}\text{C}\text{O}}}{{F}_{\text{A}\text{C}\text{O}}}\cdot\frac{{F}_{\text{A}\text{C}\text{H}4}}{{F}_{\text{I}\text{C}\text{H}4}}\right)\): logarithmic function of \(\:\frac{{F}_{\text{I}\text{C}\text{O}}}{{F}_{\text{A}\text{C}\text{O}}}\cdot\frac{{F}_{\text{A}\text{C}\text{H}4}}{{F}_{\text{I}\text{C}\text{H}4}}\). Pc: the concordance correlation coefficient (CCC). Pc (95% CI): the 95% confidence interval of CCC. Strength of agreement: the rank of the concordance correlation coefficient between the results of the two datasets.

In accordance with the 2017 ERS/ATS technical standards for lung diffusion function testing, the identification error for the concentration of the standard inhaled test gases by the lung function device must not exceed 1% [10]. In this study, both lung function devices were configured to check the inhaled gases CO and CH4 at a concentration of 0.3%, therefore the passing concentration range is between 0.297% and 0.303%. The distribution of the test gas results for CO and CH4 from the two lung function devices, and the statistical scatter plots can be seen in Table 1; Fig. 3.

Fig. 3
figure 3

Comparison of raw data distribution for FICO and FICH4 between MeAir and MasterScreen in the DLCO tests(A: FICO,B: FICH4)

Table 3 The data distribution of gas CO and CH4 of the simulator in meair and masterscreen

Simulator testing

Among the various observational indicators in the subjects, the main difference lies in the ability to recognize gas concentrations. Therefore, in the simulator test, we focused on the identification capability for checking gas concentrations, specifically analyzing and comparing FACO and FACH4.

All DLCO test simulations were conducted using the Single Breath Diffusing Capacity of the Lung for Carbon Monoxide (SB DLCO) method. We used the same simulator (Hans Rudolph, Kansas City, MO) for both lung function devices. The specific operation involved connecting the simulator to the mouthpiece of the device under test and using two precision syringes in conjunction with precise gases to simulate a single breath DLCO check operation.

In the simulator test, to compare the identification capability of the check gases at different volumes between the two sets of lung function devices, and the main inspiratory volume in the subjects is mainly between 2.0 L and 3.0 L, we used two different volumes (2.0 L, 3.0 L) with three concentrations of gases to test the MeAir and MasterScreen lung function devices. The standard concentrations of the three check gas groups were as follows: the low concentration group (c(CO%): 0.079%, c(CH4): 0.218%), the medium concentration group (c(CO%): 0.100%, c(CH4): 0.201%), and the high concentration group (c(CO%): 0.130%, c(CH4): 0.159%)). The specific execution method was to perform four simulations at each gas concentration (a total of 12 times for each device); the first simulation at each gas concentration for the device was discarded to ensure that each device system had been completely flushed [3], so the number of tests for each concentration was 3 times. At the same time, before conducting tests at different concentrations, we calibrated the machine for environmental coefficients and the gas concentrations to ensure that each test could obtain results that truly reflected the condition of the machine. The acceptable standard for the simulator gas concentration test was that the measured FACO and FACH4 had an average relative error within ± 10.0% of the standard gas concentration [17]. The distribution of the simulator test results can be seen in Table 3.

Table 4 The degree of pulmonary dispersion dysfunction in meair and masterscreen of interstitial lung disease and COPD lung disease and health

Diffusion function grade of the disease and health population groups subjects

The distribution of the grading results for the extent of lung diffusion dysfunction among the observed indicators in the ILD group and the COPD group and the health group, as well as the statistical analysis, can be found in Tables 4 and 5. The ICC values for the grading of lung diffusion dysfunction in these groups are 0.940 and 0.975 and 0.931, respectively, the above results were statistically significant (P < 0.05).

Table 5 ICC analysis of the degree of pulmonary dispersion dysfunction in interstitial lung and COPD lung disease groups
Table 6 Descriptive analysis and concordance correlation coefficient of primary and secondary end-points between masterscreen and meair

Discussion

This randomized crossover study aims to evaluate the reliability of a new structural diffusion function testing device in measuring lung diffusion function test results and assessing dysfunction grades in both respiratory disease patients and healthy populations by directly comparing two commercially available devices. Bland-Altman 95% confidence interval analysis showed good agreement in the primary indicator DLCO test results between the MeAir and MasterScreen devices, and CCC analysis also indicated a high level of agreement (CCC = 0.988). Based on the excellent agreement of DLCO and the moderate agreement of the related indicator VA, the DLCO /VA results also demonstrated moderate agreement. All statistical analysis results were statistically significant (P < 0.05). Additionally, ICC analysis revealed that the two devices also exhibited a high level of absolute agreement in assessing lung diffusion function grades (COPD group ICC = 0.940, ILD group ICC = 0.975, health group = 0.931).

Due to the poor consistency correlation of tBH, VA, FICO, FICH4, FACO and FACH4 (Table 6), any differences in these indicators could affect the numerical results of the lung diffusion function DLCO.

Therefore, we considered the possibility of special circumstances arising from internal factor differences that may offset each other. We need to conduct an in-depth analysis of the data of each factor in the calculation process, including the calculation combinations between factors (Table 6). We found that these differences offset each other in the process of calculating the diffusion function results and the results show good consistency. This phenomenon of internal differences offsetting each other has also been found in previous instrument studies [14, 18].

Analysis of the differences in CO and CH4 concentrations

From the statistical analysis of the two inhaled test gases (Tables 1, 3, and 6), it is apparent that the consistent correlation between MeAir and MasterScreen regarding the concentrations of CO and CH4 in the test gases is not satisfactory. The scatter plots (Figs. 2 and 3) and the concentration data distribution tables (Tables 1 and 3) of the results from a sample of 160 subjects indicate that both pulmonary function devices experienced fluctuations in the detection results of inhaled gas concentrations during continuous examination of a large population. According to the 2017 ERS/ATS technical standards for lung diffusion function testing, the pass rates for MeAir’s analysis of FICO and FICH4 are 63.13% and 66.88%, respectively; while the pass rates for MasterScreen’s gas analysis are 35.63% and 43.13%, respectively (Table 3) [10]. This means that MeAir’s detection of the concentration of inhaled standard gases is superior to MasterScreen, and the precision is better.

Furthermore, in the simulator test results, we can ascertain that in the detection of the gas CO, the MasterScreen device’s margin of error fluctuates between − 2.53% and 0.00%, while the MeAir device’s margin of error fluctuates between − 3.80% and − 1.00%. Regarding the detection of the CH4, the MasterScreen device’s margin of error fluctuates between − 1.83% and − 0.63%, whereas the MeAir device’s margin of error fluctuates between − 0.50% and 1.00%. Overall, this indicates that the MasterScreen is more proficient in analyzing CO compared to the MeAir, whereas the MeAir exhibits a stronger capability in recognizing the concentration of CH4 than the MasterScreen.

In summary, regarding the ability to recognize gas concentrations, the MasterScreen lung function instrument is more precise in analyzing CO concentrations, while the MeAir is better at analyzing CH4 concentrations. In the context of extensive examination scenarios, the MeAir is more precise and accurate overall compared to the MasterScreen. Given that both lung function devices are calibrated daily for environmental conditions, inhaled gas concentrations, and flow rates and speeds before conducting examinations, and only proceed with diffusion function testing on subjects after passing quality checks. We can draw a conclusion that the valve-turning structure of the MeAir does indeed ensure the stability of inhaled gas concentrations.

Analysis of the differences in alveolar volume (V A)

From the VA calculation formula, we understand that the factors involved include Vdequip, Vdanat, and the concentration ratio of the tracer gas methane (FACH4/FICH4), VI, and environmental coefficients. Since both lung function devices were calibrated for environmental factors before the examination, the environmental factors are consistent. In this project’s research, we used the FOWLER method to calculate the dead space volume, and the results for Vdanat were the same.

In terms of Vdequip, as it serves as the system dead space, it should be set constant by the manufacturer. We found that the Vdequip in the MeAir test results is consistently 283 ml, but the Vdequip results from the MasterScreen show an unstable situation, fluctuating between 142 ml and 172 ml.

Regarding the VI, the MeAir operating system is equipped with a visual exhalation platform assist line, and its diffusion test module is equipped with a new valve-turning structure, which allows the MeAir system to more accurately ensure that the subject exhales to the residual volume position. In the comparison of the observed indicator VI data results, the overall mean of MeAir is 0.99 L larger than that of MasterScreen. The results prove that the new module and exhalation platform assist line effectively standardize the operation of the diffusion function test.

Additionally, through our previous analysis of the differences in CH4 concentration between the two devices (Table 6), the consistency correlation of FACH4/FICH4 for both MasterScreen and MeAir is also poor. Combined with the results of methane concentration detection from the simulator, it indicates that MeAir has a stronger ability to recognize CH4 concentration than MasterScreen.

Therefore, even though the current consistency correlation strength of VA between MeAir and MasterScreen is moderate, we consider the results from MeAir to be more credible.

Limitations of the study

We acknowledge that there are certain limitations to this study. Firstly, in our experimental design, our research was conducted as a single-center study, and only one device each of MeAir and MasterScreen was used, which limits our ability to comprehensively evaluate the consistency of these two series of devices in diffusion function testing. Future research plans will involve multi-center studies, expand the sample size, and increase the number of MeAir and MasterScreen devices to further explore the differences between the two.

Secondly, the subjects of this study were limited to adult patients with COPD and ILD, as well as healthy adults, and did not cover other respiratory diseases or the pediatric population. In subsequent studies, we will broaden the sample range to include subjects from more age groups and disease types to further verify the wide applicability of the MeAir pulmonary function device.

Conclusions

The comprehensive analysis indicates that the MeAir lung function device and the MasterScreen lung function device both demonstrate good consistency in the measurement of key primary observation indicators for lung diffusion function tests and in the clinical assessment of the extent of lung diffusion dysfunction. It is particularly noteworthy that the MeAir lung function device shows superior stability when dealing with large populations and prolonged examination periods.

Therefore, based on the above circumstances, we believe that the diffusion function measurement system of the MeAir lung function device has met the standards for clinical application, possesses a high degree of reliability, and is suitable for widespread promotion and use.

Data availability

The pulmonary function test data and clinical information in the current study are not publicly available due to patient privacy obligations but are available from the corresponding authors on reasonable request.

Abbreviations

ERS:

European Respiratory Society

ATS:

American Thoracic Society

DLCO :

Diffusing capacity for carbon monoxide

DLCO %pred:

Diffusion capacity of carbon monoxide as a percentage of the predicted value

VA :

Alveolar volume

DLCO / VA :

DLCO to alveolar volume ratio

VI :

Inspired volume

VC:

Vital capacity

Vdequip:

Volume of equipment dead

Vdanat:

Volume of anatomical dead

tBH:

Breath-hold time

FICO :

Fractional concentration of inhaled carbon monoxide

FICH4 :

Fractional concentration of inhaled methane

FACO :

Fractional concentration of carbon monoxide in the alveolar space

FACH4 :

Fractional concentration of methane in the alveolar space

CO:

Carbon monoxide

CH4:

Methane

He:

Helium

CV%:

Coefficient of variation

T:

Ambient temperature

PB:

Barometric pressure

PH2O:

Partial pressure of water vapor

CCC:

Concordance correlation coefficient

ICC:

Intraclass correlation coefficient

References

  1. Vardar SA, Altun GD, Günerbuyuk C et al. Melatonin administration acutely decreases the diffusing capacity of carbon monoxide in human lungs. Respiration; International Review of Thoracic Diseases. 2006; 73(4): 509–513. https://doiorg.publicaciones.saludcastillayleon.es/10.1159/000088686

  2. Leech JA, Martz L, Liben A, et al. Diffusing capacity for carbon monoxide. The effects of different derivations of breathhold time and alveolar volume and of carbon monoxide back pressure on calculated results. Am Rev Respir Dis. 1985;132(5):1127–9. https://doiorg.publicaciones.saludcastillayleon.es/10.1164/arrd.1985.132.5.1127.

    Article  CAS  PubMed  Google Scholar 

  3. Jensen R, Leyk M, Crapo R, et al. Quality control of DL,CO instruments in global clinical trials. Eur Respir J. 2009;33(4):828–34. https://doiorg.publicaciones.saludcastillayleon.es/10.1183/09031936.00091208.

    Article  CAS  PubMed  Google Scholar 

  4. Neder JA, Berton DC, Muller PT, et al. Incorporating lung diffusing capacity for carbon monoxide in clinical decision making in chest Medicine. Clin Chest Med. 2019;40(2):285–305. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.ccm.2019.02.005.

    Article  PubMed  Google Scholar 

  5. Neder JA, Berton DC, O’Donnell DE. The lung function laboratory to assist clinical Decision-making in pulmonology: evolving challenges to an old Issue. Chest. 2020;158(4):1629–43. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.chest.2020.04.064.

    Article  PubMed  Google Scholar 

  6. Enright MDP. Office-based DLCO tests help pulmonologists to make important clinical decisions. Respiratory Invest. 2016;54(5):305–11. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.resinv.2016.03.006.

    Article  Google Scholar 

  7. Martinez FJ, Agusti A, Celli BR, et al. Treatment trials in young patients with chronic obstructive pulmonary disease and Pre-Chronic obstructive pulmonary disease patients: time to move Forward. Am J Respir Crit Care Med. 2022;205(3):275–87. https://doiorg.publicaciones.saludcastillayleon.es/10.1164/rccm.202107-1663SO.

    Article  PubMed  Google Scholar 

  8. Kraemer R, Gardin F, Smith HJ, et al. Functional predictors discriminating Asthma–COPD overlap (ACO) from chronic obstructive pulmonary disease (COPD). Int J Chronic Obstr Pulm Dis. 2022;17:2723–43. https://doiorg.publicaciones.saludcastillayleon.es/10.2147/COPD.S382761.

    Article  Google Scholar 

  9. Shen L, Zhang Y, Su Y, et al. New pulmonary rehabilitation exercise for pulmonary fibrosis to improve the pulmonary function and quality of life of patients with idiopathic pulmonary fibrosis: a randomized control trial. Annals Palliat Med. 2021;10(7):7289–97. https://doiorg.publicaciones.saludcastillayleon.es/10.21037/apm-21-71.

    Article  Google Scholar 

  10. Graham BL, Brusasco V, Burgos F, et al. 2017 ERS/ATS standards for single-breath carbon monoxide uptake in the lung. Eur Respir J. 2017;49(1):1600016. https://doiorg.publicaciones.saludcastillayleon.es/10.1183/13993003.00016-2016.

    Article  PubMed  Google Scholar 

  11. Madsen AC, Thomsen RS, Nymand SB, et al. Pulmonary diffusing capacity to nitric oxide and carbon monoxide during exercise and in the supine position: a test-retest reliability study. Exp Physiol. 2023;108(2):307–17. https://doiorg.publicaciones.saludcastillayleon.es/10.1113/EP090883.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Coffman KE, Carlson AR, Miller AD et al. The effect of aging and cardiorespiratory fitness on the lung diffusing capacity response to exercise in healthy humans. Journal of Applied Physiology (Bethesda, Md.: 1985), 2017, 122(6): 1425–1434. https://doiorg.publicaciones.saludcastillayleon.es/10.1152/japplphysiol.00694.2016

  13. Blakemore WS, Forster RE, Morton JW, et al. A standardized breath holding technique for the clinical measurement of the diffusing capacity of the lung for carbon monoxide. J Clin Investig. 1957;36(1 Part 1):1–17. https://doiorg.publicaciones.saludcastillayleon.es/10.1172/JCI103402.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Radtke T, DE Groot Q, Haile SR, et al. Lung diffusing capacity for nitric oxide measured by two commercial devices: a randomised crossover comparison in healthy adults. ERJ Open Res. 2021;7(3):00193–2021. https://doiorg.publicaciones.saludcastillayleon.es/10.1183/23120541.00193-2021.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Akoglu H. User’s guide to correlation coefficients. Turkish J Emerg Med. 2018;18(3):91–3. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.tjem.2018.08.001.

    Article  Google Scholar 

  16. Koo TK, LI MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability Research. J Chiropr Med. 2016;15(2):155–63. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.jcm.2016.02.012.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Hegewald MJ, MARKEWITZ B A, WILSON E L, et al. Single-breath diffusing capacity for carbon monoxide instrument accuracy across 3 health systems. Respir Care. 2015;60(3):430–6. https://doiorg.publicaciones.saludcastillayleon.es/10.4187/respcare.03512.

    Article  PubMed  Google Scholar 

  18. Matsuki T, YANAGI H, KOBA T, et al. Comparing the MiniBox™ and the Chestac-8900® for pulmonary function testing. Int J Tuberculosis Lung Disease: Official J Int Union against Tuberculosis Lung Disease. 2023;27(9):709–11. https://doiorg.publicaciones.saludcastillayleon.es/10.5588/ijtld.23.0212.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was Supported by R&D Program of Guangzhou National Laboratory (Grant No. SRPG22-018).

Author information

Authors and Affiliations

Authors

Contributions

Shihua Yao, Jinping Zheng, Yi Gao, and Yanqing Xie were responsible for proposing the research concept and designing the study protocol, including establishing observational indicators, setting inclusion and exclusion criteria, and determining the methods for group allocation. Shihua Yao was in charge of communicating with the participants and obtaining informed consent. Kuiqing Lin, Zhongping Wu, Shubing Chen, Yanqing Xie, Beilan Shen, Liping Zhong, Jiaying An, Xudong Wang, Wenting Liu, and Xinxin Yu were responsible for the study implementation, including diffusion capacity tests and participant screening. Shihua Yao handled data collection, acquisition, and cleaning. Shihua Yao and Tuping Fu conducted the statistical analysis. Shihua Yao drafted the manuscript, while Shihua Yao, Jinping Zheng, and Yi Gao revised the final version and are accountable for the accuracy and integrity of the work.

Corresponding authors

Correspondence to Jinping Zheng or Yi Gao.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Ethics Committee of the First Affiliated Hospital of Guangzhou Medical University (ES-2023-126-02), and all participants were informed of the details of the trial and have signed informed consent forms before the experiment.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yao, S., Fu, T., Lin, K. et al. The Reliability of a Novel Structured Testing Device for Single-breath Lung Carbon Monoxide Uptake: A Randomized Comparison Crossover Study. BMC Pulm Med 25, 230 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12890-025-03571-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12890-025-03571-7

Keywords