Diagnostic performance of a doppler radar-based sleep apnoea testing device

Abstract

Background

Inpatient polysomnography (PSG) is the gold standard for the diagnosis of obstructive sleep apnoea (OSA), however, both complexity and costs limit the availability of this examination. Home sleep apnoea testing devices are a diagnostic alternative in patients with increased risk of OSA. We evaluated the diagnostic performance of a Doppler radar technology based, contactless sleep apnoea testing device (CSATD) in a cohort of patients with a clinically increased risk of OSA.

Methods

Monocentric prospective study. Sleep monitoring with the CSATD SleepizOne + without pulse oximetry (Sleepiz AG, Switzerland) was performed simultaneously with elective inpatient PSG. PSG was analysed blinded to the CSATD results and according to AASM 2012 criteria by certified sleep physicians. The CSATD data were analysed automatically and independently by a dedicated software.

Results

A total of 102 patients, 60.8% male, with an average age of 55 ± 15 years and body mass index of 30 ± 6 kg/m2 were included in the analysis. The sensitivity and specificity of the CSATD for a PSG apnoea-hypopnoea-index (AHI) of ≥ 5/h were 0.89 (95%CI: 0.83–0.96) and 0.88 (95%CI: 0.73-1.0). The negative and positive predictive values were 0.62 (95%CI: 0.42–0.82) and 0.97 (95%CI: 0.94-1.0). The diagnostic agreement for the diagnosis of OSA (defined as PSG AHI ≥ 5/h) was 89.8% and 100% using a CSATD AHI threshold of ≥ 5/h (n = 79/88) and ≥ 15/h (n = 61/61). However, the concordance was poor in the classification of OSA severity, with 50% (13/26) concordance for mild, 38% (10/26) for moderate, and 76% (25/33) for severe OSA respectively.

Conclusion

CSATD accurately identifies patients with OSA, particularly using an AHI threshold of ≥ 15/h. However, it performs subpar in disease severity stratification.

Clinical trial registration

This trial was registered on the International Clinical Trials Registry Platform, ISRCTN45778591.

Peer Review reports

Background

Symptomatic obstructive sleep apnoea (OSA) is common, with a prevalence of 5% in women and up to 14% in men [1]. Left untreated, OSA often impairs quality of life [2] and is associated with an increased cardiovascular risk [3, 4]. If OSA is suspected, sleep evaluation and appropriate diagnostic testing is recommended [5].

Although, in-laboratory overnight polysomnography (PSG) is widely accepted as the reference standard for the diagnosis of OSA, both its complexity and costs might limit access to PSG [5]. The American Academy of Sleep Medicine (AASM) has approved the use of home sleep apnoea testing (HSAT) in selected adult patients, mainly defined as patients with clinically increased risk for moderate to severe OSA and without risk factors for non-obstructive sleep-disordered breathing [5, 6]. HSAT devices offer many advantages, including patient comfort, lower costs, better availability, and prompt analysis [7, 8]. To date, the majority of single or multi-channel devices are worn by the patient and measure parameters such as airflow, (chest) movement, oxygen saturation and/or peripheral arterial tone [5, 9]. Additionally, few contactless sleep apnoea testing devices (CSATD) have been investigated [10]. The main categories of non-contact breathing and sleep monitoring are based on visual [11], audio [12] or radiofrequency technologies [13,14,15,16].

SleepizOne + is a new CSATD based on Doppler radar technology [13, 17, 18]. Electromagnetic waves emitted by the device are reflected by the sleeping patients and then analysed, enabling the detection of subtle chest movements. This technique can be used to determine both respiratory rate and heart rate [13, 18].

The objective of this study was to analyse the diagnostic performance of the CSATD (SleepizOne+) in detecting sleep apnoea in a clinical cohort of 102 patients with suspected OSA who underwent in-laboratory PSG.

Methods

This prospective, monocentric study was performed at the Clinic of Respiratory Medicine and Pulmonary Cell Research at the University Hospital of Basel, Switzerland. The Ethics Committee northwest/central Switzerland approved the study (EKNZ 2018–02086). It was carried out according to the Declaration of Helsinki and Good Clinical Practice guidelines. Patients undergoing in-laboratory PSG at the University Hospital of Basel between May 2020 and January 2021 were approached for study inclusion and only admitted if they gave their informed consent.

The PSGs were analysed by trained sleep physicians blinded to the CSATD results according to the AASM 2012 criteria [19]. The CSATD (SleepizOne+) data were analysed independently and externally by the company (Sleepiz AG, Switzerland), blinded to the PSG results and the patients’ personal data. The CSATD was installed in close proximity to the patient (approximately 40–50 cm) in addition to the standard PSG equipment. It emits an electromagnetic signal at a fixed frequency of 24 GHz which is reflected from the patient’s surface, whereby the duvet does not cause relevant signal interference [13]. The reflected signal is received by the device’s transceiver and processed. Thoracic and abdominal movements, such as breathing, cause small relative distance changes and can be deduced from the reflected signals [13, 18]. The data were analysed by the company’s software after they have been uploaded to a dedicated server. In the present study, the CSATD was not connected to a pulse oximeter. Photographs of the installed device and the reported signal can be found in the supplementary appendix, further information regarding the mechanism of the CSATD can be found in the literature [13, 18].

Proper HSAT diagnostic testing requires at least 4 h of adequate recording [5]. In this study, as shown in Fig. 1, we excluded sleep studies with an estimated total sleep time (TST) of less than 4 h (n = 17). Sleep studies under positive airway pressure therapy were also excluded (n = 1). Recording was started before the patients went to bed. The total bed time estimate is the period of time during which the device detects that a patient is lying in front of it. The TST is the estimated time spent sleeping within the recording period. All CSATD parameters, including apnoea-hypopnoea-index (AHI), were calculated automatically by the software.

The term sleep apnoea was defined on the basis of an AHI cut-off value of ≥ 5 events/h and classified as mild (5 ≤ AHI < 15/h), moderate (15 ≤ AHI < 30/h) or severe (AHI ≥ 30/h) [5]. The oxygen desaturation cut-off of ≥ 3% was used to determine oxygen desaturation index (ODI) or hypopnoea. It is important to note that the CSATD did not differentiate between central and obstructive apnoea events. Given that only one patient was diagnosed with central sleep apnoea (CSA), the term OSA was used for simplicity and readability. An Epworth Sleepiness Scale (ESS) score of ≥ 11/24 points was used as cut-off for excessive daytime sleepiness [20].

The data, including patient demographics and medical history, were obtained from electronic medical records of the out- and inpatient departments of the University Hospital of Basel. Categorical parameters were summarized as counts and percentages, continuous parameters as means and standard deviation (SD). The Lin’s concordance coefficient and Bland-Altman diagrams were used to examine the agreement between the CSATD and PSG measurements. To evaluate the diagnostic value of the tested method (CSATD) to predict sleep apnoea, receiver operating characteristic curves were drawn. The rate of true positive predictions at different cut-off points were plotted versus the rate of false positive predictions. The area under the curve is reported and the cut-off point with the maximum Youden’s index was chosen to determine the associated sensitivity and specificity of the test. The sleep parameters determined using both instruments were compared using the Wilcoxon’s signed rank test. Generalized linear regression models were performed to determine the effect of body position on the difference between PSG AHI and CSATD AHI. An alpha level of 0.05 was set as the limit of statistical significance. IBM SPSS Statistics for Windows, version 25 (IBM Corp., Armonok N.Y., USA) and SAS version 9.4 (SAS Institute Inc., Cary, NC, USA) were used for the statistical analyses.

Results

Between May 2020 and January 2021, a total of 120 patients underwent PSG with simultaneous CSATD, of whom 102 were included in the final analysis (Fig. 1).

Fig. 1
figure 1

Schematic representation of patient inclusion in the study. Polysomnography and the contactless sleep apnoea testing device (CSATD) were performed simultaneously. CPAP: Continuous positive airway pressure

The patient population was predominantly male (60.8%). The average age was 55.3 ± 15.3 years and the mean body mass index (BMI) was 30.1 ± 6.4 kg/m2. Daytime sleepiness, defined by an ESS ≥ 11/24 points, was present in 32.4% (23/71) of the subjects. The most prevalent comorbidities were arterial hypertension (52.0%), chronic kidney disease (CKD 25.5%), depression (17.7%) and diabetes mellitus (16.6%) (Table 1).

Table 1 Sociodemographic and clinical characteristics

OSA, as diagnosed by PSG, was highly prevalent in this cohort (85/102, 83.3%) (Table 2). The prevalence of mild OSA was 25.5% (26/102) and that of moderate to severe OSA was 57.8% (59/102). The CSATD had an 89.8% agreement with the PSG for the overall diagnosis of OSA (79/88). In all patients with moderate to severe OSA in the CSATD (AHI ≥ 15/h) the diagnosis of sleep apnoea, defined as PSG AHI ≥ 5/h, was confirmed (61/61). Moderate OSA was found in 17.0% (7/41) and severe OSA in 2.4% (1/41) of patients with no or mild OSA in the CSATD.

Table 2 Diagnostic performance by OSA severity

The area under the receiver operating characteristic curves (ROC AUC) (Suppl. Figure 3) was used to show the diagnostic performance of the CSATD compared to PSG for three different AHI cut-off points. The AUC values for an AHI ≥ 5/h, ≥ 15/h or ≥ 30/h, were 0.92, 0.87 and 0.89, respectively. Table 3 summarizes the performance in terms of sensitivity, specificity, negative and positive predictive values applying these diagnostic thresholds. The sensitivity and specificity of the CSATD for OSA diagnosis (AHI ≥ 5/h) was 89.4% and 88.2% (Table 3).

Table 3 Diagnostic performance of the CSATD in the study population

The concordance correlation coefficient for AHI measured by PSG and the CSATD was 0.78 indicating a good agreement between both sleep studies (Fig. 2). In the Bland-Altman diagram, we showed that 5.9% (n = 6/102) of the data points fall outside the limits of agreement (+ 1.96 SD and − 1.96 SD).

Fig. 2
figure 2

Concordance between AHI measured by the CSATD and polysomnography (a) Scatter plot and (b) Bland Altman diagram. AHI: Apnoea-hypopnoea-index CSATD: Contactless sleep apnoea testing device. PSG: Polysomnography

When analysing the concordance with PSG in classifying the severity of OSA, an agreement of 50% (13/26) for mild, 38% (10/26) for moderate, and 76% (25/33) for severe OSA was found (Fig. 3). The ROC AUC analysis showing the performance of the CSATD in classifying OSA severity is provided in Suppl. Figure 4. The AUC values for no sleep apnoea, mild, moderate and severe OSA were 0.92, 0.67, 0.47 and 0.89. Post-hoc analysis of the PSGs of patients with markedly discrepant results, CSATD AHI < 5/h but moderate to severe OSA on PSG, showed that these patients (n = 3) had hypopnoea-predominant OSA with an elevated ODI, but an apnoea index of ≤ 6.5/h.

Fig. 3
figure 3

Sankey plot showing sleep apnoea severity classification in both sleep studies (CSATD and polysomnography (PSG)). CSATD: Contactless sleep apnoea testing device

The sleep variables determined with the CSATD and PSG are shown in Suppl. Table 2. A significantly longer mean TST was recorded with the CSATD than with the PSG (368.7 ± 5.8 min vs. 344.9 ± 7.8 min, p = 0.006). Mean AHI (24.8 ± 1.8 /h vs. 24.9 ± 2.2 /h) and heart rate (64.9 ± 0.9 bpm vs. 64.5 ± 0.9 bpm, p = 0.77) were similar using both diagnostic tools. Suppl. Table 3 provides an overview of the sleep variables that are measured exclusively with PSG. An analysis of these parameters, with a particular focus on sleep-disordered breathing, showed a mean hypopnoea index of 13.0 ± 3.0 /h, an ODI of 24.0 ± 2.1 /h and a mean oxygen saturation of 92.3 ± 0.3%. The majority of patients showed a position-dependent aggravation of AHI. The mean AHI in the supine position was with 36.2 ± 3.1/h markedly higher than the overall mean AHI of 24.9 ± 2.2 /h. In regard to sleep position in percentage of TST, the supine position was predominant with a mean of 53.1 ± 3.0%. The lateral sleeping position was found in an average of 45.7 ± 2.3% of the TST.

Generalized linear regression models were performed to determine the effect of body position on the difference between PSG AHI and CSATD AHI (Suppl. Table 4). Especially the supine position led to an increased AHI difference between the two sleep studies. An increase of 1% in supine position (of TST), increased the difference in AHI by 0.12/h, p = 0.0090. It is important to note that the difference also increased with increasing PSG AHI (by 0.33/h per unit increase in PSG AHI, data not shown). When analysing the influence of other parameters on the AHI difference, the hypopnoea index in particular was associated with an increased AHI difference between the two sleep studies. An increase in the hypopnoea index by 1/h increased the AHI difference by 0.75/h (p < 0.0001).

Discussion

The present study analysed the accuracy of the SleepizOne + device, a novel CSATD, for the diagnosis of sleep apnoea in a representative clinical cohort of patients with suspected OSA, in comparison to PSG. We found good agreement in the diagnosis of OSA, especially when using a CSATD AHI cut-off of ≥ 15/h, but discordance in OSA severity stratification or sleep time estimation.

Non-contact sleep monitoring devices are an emerging technology for the assessment of OSA [10, 11, 14,15,16]. SleepWise is based on automatic video analysis for sleep apnoea diagnosis and showed a sensitivity of 100% and a specificity of 83% for an AHI of ≥ 5/h in a trial including 50 patients [11]. Another device, OrbSense, monitors sleep related breathing using microwave radar technology [14, 16]. Similarly, OrbSense had a 96% sensitivity, but 56% specificity in detecting sleep apnoea compared to PSG. The severity classification was concordant in 69.6% of examinations [16]. SleepMinder, a low-power radiofrequency energy device, showed a sensitivity and specificity of 98% and 47%, respectively, using a threshold AHI of ≥ 5/h, whereas the AUC was 0.97 for an AHI ≥ 15/h [15].

Like other non-contact devices, the CSATD analysed in this study showed good diagnostic performance with a sensitivity of 89.4% (95% CI 82.9–96.0), a specificity of 88.2% (95% CI 72.9–100%) and an AUC of 0.92 (95% CI 0.87–0.97) for the diagnosis of sleep apnoea defined by PSG AHI ≥ 5/h. When CSATD AHI is ≥ 15/h, sleep apnoea diagnosis of any severity is likely as it was herein confirmed by PSG in all cases (61/61). Gross-Isselmann et al. also analysed the performance of SleepizOne + compared to PSG in a slightly smaller cohort of patients [18]. They used a binary classification of sleep apnoea with a cut-off AHI of ≥ 15/h to differentiate between moderate to severe sleep apnoea and mild or no sleep apnoea. Using that cut-off, they found a sensitivity of 85.4% and a specificity of 88.1%, which are comparable to our results applying the same cut-off with a sensitivity of 83.1% (CI 73.5–92.6) and specificity of 83.7% (CI 72.7–94.8) [18]. By including data from a pulse oximeter, the device was no longer completely contactless in this study, but showed an increase in sensitivity to 87.8% and specificity to 98.3% [18].

In addition to improved precision, peripheral arterial oxygen saturation measurements provide clinicians with further clinically relevant information, as intermittent oxygen desaturations have been independently associated with cardiovascular disease [21]. In our study we observed three patients without sleep apnoea in the CSATD but moderate to severe sleep apnoea in PSG. These patients showed hypopnoea-dominant sleep apnoea with an increased ODI but an almost normal apnoea index (≤ 6.5/h). Since an increase in the hypopnoea index by 1/h in the PSG increased the AHI difference between PSG and CSATD by 0.75/h, recording the ODI could reduce such discrepant findings or at least alert the physician that sleep-disordered breathing is to be expected.

For uncomplicated patients at increased risk of moderate to severe OSA, the AASM practice guidelines state that OSA can be diagnosed using HSAT devices, which have been shown to have similar sensitivity, specificity and AUC to the CSATD investigated in this study [5]. The potential benefits of HSAT include improved patient comfort, increased availability, and reduced costs [5]. Our data suggest that the diagnosis of OSA can be assumed in patients with a high pre-test probability and moderate to severe findings in the CSATD. However, the low negative predictive value of the CSATD in our population suggests that it may not be sufficient to rule out OSA.

The overall good performance of the device in diagnosing sleep apnoea was tempered by its limited ability to correctly classify the severity of OSA according to the current reference standard, polysomnography (Fig. 3). This was also reflected in a wider limit of agreement in the Bland-Altman diagram (Fig. 2). From a clinical perspective, this represents a considerable limitation of the device. Accurate classification of OSA severity is of clinical relevance as it is important for treatment decisions. In contrast to mild OSA, (moderate to) severe OSA has been shown to be a significant cardiovascular risk factor for patients [22, 23].

Caution is required in patients with suspected or increased risk of CSA, as this device did not differentiate between obstructive and central apnoeas. The effect of CSA on diagnostic accuracy could not be systematically analysed in this study. One patient in this cohort had predominant CSA and was classified as having severe sleep apnoea in both sleep studies.

An important but inherent limitation of our study was the laboratory testing environment, which affected both the installation of the CSATD devices and the patients’ sleep quality and efficiency. It can be assumed that the body position was also influenced by the setting, as the supine position predominated with a mean value of 53.1 ± 3.0% of the TST [24].

Although this is the largest study to date comparing the CSATD with PSG, the generalisability of the results is limited by the single-centre design and the sample size, resulting in a low statistical power of the study (McNemar test, comparison of both sleep studies for OSA diagnosis (Suppl. Figure 5)).

The AASM practice guidelines recommended high pre-test probabilities when a HSAT is performed to diagnose sleep apnoea [5]. A further limitation of our data is that, although the pre-test probability for OSA was clinically assessed, it was not documented in a standardised way. Scores such as the STOP-BANG [25], Berlin questionnaire [26] or NoSAS [27], which are important instruments for assessing the pre-test probability for OSA, could therefore not be provided. The ESS was documented in 69% of the patients. However, the clinical assessment and pre-selection of the patients, some of whom had already undergone HSAT screening tests before being referred for PSG, resulted in a high prevalence of OSA (83.3%) in our clinical population.

Conclusions

The CSATD performed well compared to PSG in the diagnosis of (obstructive) sleep apnoea in selected patients with clinical suspicion of OSA. Further development, focusing on better sleep severity stratification and differentiation between central and obstructive events, might lead to increased diagnostic options for clinicians in an out-patient setting.

Data availability

Data is provided within the manuscript or supplementary information files. The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AASM:

American Academy of Sleep Medicine

AHI:

Apnoea-hypopnoea-index

AUC:

Area under the curve

BMI:

Body mass index

CI:

Confidence interval

CPAP:

Continuous positive airway pressure

CSA:

Central sleep apnoea

CSATD:

Contactless sleep apnoea testing device

ESS:

Epworth Sleepiness Scale

HSAT:

Home sleep apnoea testing

ODI:

Oxygen desaturation index

OSA:

Obstructive sleep apnoea

PSG:

Polysomnography

REM:

Rapid eye movement

ROC:

Receiver operating characteristic

SD:

Standard deviation

TST:

Total sleep time

References

  1. Peppard PE, Young T, Barnet JH, Palta M, Hagen EW, Hla KM. Increased prevalence of sleep-disordered breathing in adults. Am J Epidemiol. 2013;177(9):1006–14.

    PubMed  PubMed Central  Google Scholar 

  2. D’Ambrosio C, Bowman T, Mohsenin V. Quality of life in patients with obstructive sleep apnea: effect of nasal continuous positive airway pressure–a prospective study. Chest. 1999;115(1):123–9.

    PubMed  Google Scholar 

  3. Gottlieb DJ, Yenokyan G, Newman AB, O’Connor GT, Punjabi NM, Quan SF, et al. Prospective study of obstructive sleep apnea and incident coronary heart disease and heart failure: the sleep heart health study. Circulation. 2010;122(4):352–60.

    PubMed  PubMed Central  Google Scholar 

  4. Patil SP, Billings ME, Bourjeily G, Collop NA, Gottlieb DJ, Johnson KG, et al. Long-term health outcomes for patients with obstructive sleep apnea: placing the agency for healthcare research and quality report in context-a multisociety commentary. J Clin Sleep Med. 2024;20(1):135–49.

    PubMed  PubMed Central  Google Scholar 

  5. Kapur VK, Auckley DH, Chowdhuri S, Kuhlmann DC, Mehra R, Ramar K, et al. Clinical practice guideline for diagnostic testing for adult obstructive sleep apnea: an American Academy of Sleep Medicine clinical Practice guideline. J Clin Sleep Med. 2017;13(3):479–504.

    PubMed  PubMed Central  Google Scholar 

  6. Rosen IM, Kirsch DB, Chervin RD, Carden KA, Ramar K, Aurora RN, et al. Clinical use of a home sleep apnea test: an American Academy of Sleep Medicine position statement. J Clin Sleep Med. 2017;13(10):1205–7.

    PubMed  PubMed Central  Google Scholar 

  7. Patel SR. Obstructive sleep apnea. Ann Intern Med. 2019;171(11):ITC81–96.

    PubMed  Google Scholar 

  8. Skomro RP, Gjevre J, Reid J, McNab B, Ghosh S, Stiles M, et al. Outcomes of home-based diagnosis and treatment of obstructive sleep apnea. Chest. 2010;138(2):257–63.

    PubMed  Google Scholar 

  9. Röcken J, Schumann DM, Herrmann MJ, Veitz S, Franchetti L, Grize L, et al. Peripheral arterial tonometry versus polysomnography in suspected obstructive sleep apnoea. Eur J Med Res. 2023;28(1):251.

    PubMed  PubMed Central  Google Scholar 

  10. Khalil C, Zarabi S, Kirkham K, Soni V, Li Q, Huszti E, et al. Validity of non-contact methods for diagnosis of obstructive sleep apnea: a systematic review and meta-analysis. J Clin Anesth. 2023;87:111087.

    PubMed  Google Scholar 

  11. Abad J, Muñoz-Ferrer A, Cervantes M, Esquinas C, Marin A, Martínez C, et al. Automatic video analysis for obstructive sleep apnea diagnosis. Sleep. 2016;39(8):1507–15.

    PubMed  PubMed Central  Google Scholar 

  12. Narayan S, Shivdare P, Niranjan T, Williams K, Freudman J, Sehra R. Noncontact identification of sleep-disturbed breathing from smartphone-recorded sounds validated by polysomnography. Sleep Breath. 2019;23(1):269–79.

    PubMed  Google Scholar 

  13. Bujan B, Fischer T, Dietz-Terjung S, Bauerfeind A, Jedrysiak P, Große Sundrup M, et al. Clinical validation of a contactless respiration rate monitor. Sci Rep. 2023;13(1):3480.

    CAS  PubMed  PubMed Central  Google Scholar 

  14. Li CX, Zhang YF, Zhu Z, Lu FY, Wang Y, Zhang LY, et al. Diagnosis of obstructive sleep apnea using a bio-radar contact-free system compared with an established HST device in older adults. Sleep Health. 2023;9(3):381–6.

    PubMed  Google Scholar 

  15. Zaffaroni A, Kent B, O’Hare E, Heneghan C, Boyle P, O’Connell G, et al. Assessment of sleep-disordered breathing using a non-contact bio-motion sensor. J Sleep Res. 2013;22(2):231–6.

    PubMed  Google Scholar 

  16. Zhao R, Xue J, Dong XS, Zhi H, Chen J, Zhao L, et al. Screening for obstructive sleep apnea using a contact-free system compared with polysomnography. J Clin Sleep Med. 2021;17(5):1075–82.

    PubMed  PubMed Central  Google Scholar 

  17. Tschopp S, Borner U, Caversaccio M, Tschopp K. Long-term night-to-night variability of sleep-disordered breathing using a radar-based home sleep apnea test: a prospective cohort study. J Clin Sleep Med. 2024;20(7):1079–86. https://doiorg.publicaciones.saludcastillayleon.es/10.5664/jcsm.11070. PMID: 38415722; PMCID: PMC11217624.

  18. Gross-Isselmann JA, Eggert T, Wildenauer A, Dietz-Terjung S, Grosse Sundrup M, Schoebel C. Validation of the sleepiz One + as a radar-based sensor for contactless diagnosis of sleep apnea. Sleep Breath. 2024;28(4):1691–9.

    PubMed  PubMed Central  Google Scholar 

  19. Berry RB, Budhiraja R, Gottlieb DJ, Gozal D, Iber C, Kapur VK, et al. Rules for scoring respiratory events in sleep: update of the 2007 AASM manual for the scoring of sleep and associated events. Deliberations of the sleep apnea definitions task force of the American academy of sleep medicine. J Clin Sleep Med. 2012;8(5):597–619.

    PubMed  PubMed Central  Google Scholar 

  20. Sánchez-de-la-Torre M, Sánchez-de-la-Torre A, Bertran S, Abad J, Duran-Cantolla J, Cabriada V, et al. Effect of obstructive sleep apnoea and its treatment with continuous positive airway pressure on the prevalence of cardiovascular events in patients with acute coronary syndrome (ISAACC study): a randomised controlled trial. Lancet Respir Med. 2020;8(4):359–67.

    PubMed  Google Scholar 

  21. Azarbarzin A, Sands SA, Stone KL, Taranto-Montemurro L, Messineo L, Terrill PI, et al. The hypoxic burden of sleep apnoea predicts cardiovascular disease-related mortality: the osteoporotic fractures in men study and the sleep heart health study. Eur Heart J. 2019;40(14):1149–57.

    PubMed  Google Scholar 

  22. Chowdhuri S, Quan SF, Almeida F, Ayappa I, Batool-Anwar S, Budhiraja R, et al. An official American Thoracic Society research statement: impact of mild obstructive sleep apnea in adults. Am J Respir Crit Care Med. 2016;193(9):e37–54.

    PubMed  Google Scholar 

  23. Yeghiazarians Y, Jneid H, Tietjens JR, Redline S, Brown DL, El-Sherif N, et al. Obstructive sleep apnea and cardiovascular disease: a scientific statement from the American Heart Association. Circulation. 2021;144(3):e56–67.

    CAS  PubMed  Google Scholar 

  24. Yo SW, Joosten SA, Wimaleswaran H, Mansfield D, Thomson L, Landry SA, et al. Body position during laboratory and home polysomnography compared to habitual sleeping position at home. J Clin Sleep Med. 2022;18(9):2103–11.

    PubMed  PubMed Central  Google Scholar 

  25. Chung F, Abdullah HR, Liao P, STOP-Bang Questionnaire. A practical approach to screen for obstructive sleep apnea. Chest. 2016;149(3):631–8.

    PubMed  Google Scholar 

  26. Netzer NC, Stoohs RA, Netzer CM, Clark K, Strohl KP. Using the Berlin questionnaire to identify patients at risk for the sleep apnea syndrome. Ann Intern Med. 1999;131(7):485–91.

    CAS  PubMed  Google Scholar 

  27. Marti-Soler H, Hirotsu C, Marques-Vidal P, Vollenweider P, Waeber G, Preisig M, et al. The NoSAS score for screening of sleep-disordered breathing: a derivation and validation study. Lancet Respir Med. 2016;4(9):742–8.

    PubMed  Google Scholar 

Download references

Funding

Open Access funding enabled and organized by Projekt DEAL.

Clinic of Respiratory Medicine and Pulmonary Cell Research, University Hospital Basel.

Author information

Authors and Affiliations

Authors

Contributions

JR contributed to data collection, ensuring accuracy of data, discussion of results, statistical analysis, writing of the manuscript, finalization of the manuscript and approval of the submitted article. CED contributed to data collection. MJH, KJ, WS, and MT contributed to data collection and clinical work. LG contributed to the statistical analysis of the data, LG and AD contributed to writing of the manuscript. DS conceived the project, contributed in clinical work, statistical analysis, writing of the manuscript, integrity, and accuracy of data, and preparation and approval of the submitted article. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Daiana Stolz.

Ethics declarations

Ethics approval and consent to participate

The Ethikkommission Beider Basel, approved the study (EKNZ 2018–02086) and the subjects provided written informed consent. This study was conducted in accordance with the amended Declaration of Helsinki.

Consent for publication

All subjects provided written informed consent.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Röcken, J., Darie, A.M., Grize, L. et al. Diagnostic performance of a doppler radar-based sleep apnoea testing device. BMC Pulm Med 25, 150 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12890-025-03618-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12890-025-03618-9

Keywords