Skip to main content

Comparison of 5% sodium bicarbonate and 10% sodium chloride as contrast agents for lung perfusion with electrical impedance tomography: a prospective clinical study

Abstract

Introduction

This study aimed to compare sodium bicarbonate (NaHCO3) and sodium chloride (NaCl) as contrast agents for regional lung perfusion imaging with electrical impedance tomography in critically ill patients with respiratory failure.

Methods

15 mL 5% NaHCO3 and 10 mL 10% NaCl were sequentially administered in all the enrolled patients with respiratory failure. EIT maps were divided into four cross-regions: upper right (UR), upper left (UL), lower right (LR), and lower left (LL).

Result

A total of 20 pair bolus injections from 16 mechanically ventilated patients were obtained. Compared to NaHCO3, NaCl caused a larger drop of maximum impedance amplitude after the bolus injection (1276 ± 329 vs. 509 ± 159 AU, P < 0.001). Regional perfusion distribution (%) of four cross-regions between two indicators were significantly correlated (R2 = 0.90 for regional perfusion%, p < 0.001; R2 = 0.93 for regional ventilation/perfusion matching, p < 0.001) in 80 pair measurements. The Bland–Altman analysis showed a strong overall agreement in regional perfusion distribution (%) [mean bias 0.09% (95% confidence interval, CI: -0.91%, 1.09%), lower limits of agreement (LOA) -8.03% (95% CI: -9.75%, -6.31%), upper LOA 8.21% (95% CI: 6.48%, 9.93%)] between indicators in 80 pair measurements.

Conclusion

EIT with NaHCO3 as a contrast agent yields a high agreement of regional lung perfusion compared to NaCl in critically ill patients with respiratory failure.

Peer Review reports

Introduction

Electrical impedance tomography (EIT) is a non-invasive and radiation-free visualizing tool for the assessment of lung ventilation and perfusion. It has enabled the bedside monitoring of lung perfusion measurement through the indicator-based method for critically ill patients in the intensive care unit (ICU) [1, 2]. In recent years, the saline bolus-based EIT method for perfusion assessment has been widely investigated [3,4,5]. The reconstructed regional ventilation-perfusion (V/Q) images have many potentials in clinical. It has been suggested that lung V/Q profiles can help to identify massive pulmonary embolism at the bedside [1]. For patients with acute respiratory distress syndrome (ARDS), it can provide an indication of disease severity and prognosis [6, 7]. The EIT V/Q image also has the potential to guide the management of mechanical ventilation in postoperative cardiac surgery patients [8, 9].

However, concerns may be raised regarding the selection of the contrast agent. Currently, hypertonic sodium chloride (NaCl) is the most commonly used contrast agent. Clinical studies have shown that hyperchloremia resulting from saline-based fluid resuscitation may contribute to acute kidney injury and adverse outcomes [10, 11]. Therefore, the potential risks associated with the use of hypertonic NaCl as a contrast agent for EIT-based lung perfusion assessment should not be ignored. These risks prohibit the frequent use of hypertonic NaCl bolus within a short time period. It is, therefore, desirable to find an alternative contrast agent that enables repeated injections in a safer and less damaging way to meet the need for continuous monitoring of pulmonary blood flow in critically ill patients.

As a commonly used therapeutic fluid in shock resuscitation, sodium bicarbonate (NaHCO3) has the potential for applications. Recently, two experimental studies indicated that NaHCO3 might be an alternative option to hypertonic NaCl [12, 13]. NaHCO3 has been found to be generally consistent with hypertonic NaCl as an indicator of lung perfusion in porcine models [13]. However, whether NaHCO3 results in acceptable bias is unknown in critically ill patients. This clinical study aims to further investigate the correlation and agreement of lung perfusion and ventilation/perfusion (V/Q) matching between these two contrast indicators in critically ill patients with respiratory failure.

Method

Study design and population

This is a single-center, prospective, cross-over study (ClinicalTrials.gov, NCT06868810). The study was approved by the Institutional Research and Ethics Committee of the Peking Union Medical College Hospital (I-24PJ0505). Written informed consent was obtained from all participants or their next of kin. Mechanically ventilated patients in the intensive care unit with lung disease were screened. They were included if a central venous catheter was placed due to routine clinical treatments and if regional V/Q information was of interest. Exclusion criteria were: contraindications to the use of EIT (automatic implantable cardioverter defibrillator, chest wounds limiting electrode belt placement, implantable pumps, etc.), severe hyperchloremia (> 155 mmol/L), unable to tolerate breath-hold maneuver.

Study protocol

Each enrolled patient has received two separate bolus injections of 15 mL of 5% NaHCO3 and 10 mL of 10% NaCl (except for one patient who received 5 pairs of injection). There was a 10-minute washout period between the two injections. The order of injections was determined randomly by flipping a coin. The procedure of bolus injection was performed based on our previous study [1]. Each injection was given via the central venous catheter during a breathing pause of at least 8 s at the end-expiratory hold. The patients’ status remained unchanged throughout the study period.

EIT methods

EIT measurements were performed with PulmoVista 500 (Dräger Medical, Lübeck, Germany). A silicone EIT belt with 16 surface electrodes was placed around the patient’s thorax at the 4th -5th intercostal space level in the supine position. Data were continuously recorded at 20 Hz. EIT data were analyzed offline using customized software programmed with MATLAB R2023 (MathWorks Inc., Natick, MA).

The functional ventilation map was reconstructed from an averaged period of tidal breathing [14]. The functional perfusion map was derived by calculating the slope of the impedance-time curve after contrast injection [15]. Regions of ventilation and perfusion were defined as pixels exceeding 20% of the maximum value of the functional ventilation and perfusion maps. Three types of regions were determined: the region that is mostly ventilated (RV), the region that is mostly perfused (RP), and the region with both ventilation and perfusion (RV+P). Hence, regional V/Q match was calculated as follows:

V/Q match % = RV+P / (RV+RP+RV+P) × 100% [16].

Outcome measurements

EIT images were divided into symmetrical, non-overlapping, four cross-regions: upper right (UR), upper left (UL), lower right (LR), and lower left (LL). The four regions of regional lung perfusion distribution (%) and regional V/Q match (%) were calculated [1]. Hence, 4 pairs of regional perfusion distribution and 4 pairs of V/Q matching were obtained in one EIT map.

Statistical analysis

The paired t-test was used to compare the change in impedance between the two indicators. Correlation and agreement between two contrast agents for regional lung perfusion and EIT-related parameters were performed by Pearson linear correlation and Bland–Altman analyses, respectively. Statistical analyses were conducted in R (Version 4.2.3) and the software package SPSS 25.0 (IBM Corp. Armonk, NY, USA). P < 0.05 denotes statistical significance.

Results

15/16 patients only received one pair, including two fluid contrast bolus for lung perfusion, and 1/16 patient received five pairs of fluid bolus on different days. As a result, a total of 20 pairs of 15 ml 5% NaHCO3 and 10 ml 10% NaCl fluid injections were recorded in the 16 patients. Moreover, one EIT image was divided into four regions. Hence, a total of 80 pairs (20 × 4) of regional perfusion and V/Q by the two solutions were obtained.

Patient characteristics

The baseline APACHE II score was 18 ± 5, and PaO2/FiO2 was 210 ± 79 mmHg. The etiology included pneumonia (9/16), extrapulmonary infection (5/16), pulmonary embolism (1/16), and alveolar hemorrhage (1/16). Patients’ demographics and clinical characteristics are summarized in Table 1.

Table 1 Patients’ demographics and clinical characteristics

Outcomes

Compared to 15 ml 5% NaHCO3, 10 ml 10% NaCl caused a larger decrease in impedance after the bolus injection (1276 ± 329 vs. 509 ± 159 arbitrary unit, P < 0.001). Figure 1 shows the impedance-time curves during tidal breathing and bolus injection periods from a single patient. Ventilation distribution, perfusion distribution, and regional V/Q match images were compared (Fig. 1A and B, 2nd– 4th columns). A similar perfusion and V/Q match images were found between the two contrast agents.

Fig. 1
figure 1

An individual patient’s data shows two contrasts agents for perfusion and regional V/Q estimation in electrical impedance tomography. 10 ml 10% NaCl caused a greater drop of impedance amplitude than 15 ml 5% NaHCO3 during the breath hold period. A similar perfusion and V/Q match images were found between the two indicators. NaHCO3, sodium bicarbonate; NaCl, sodium chloride

Regional perfusion distribution (%) and V/Q match (%) in four cross-regions (UR, UL, LR, and LL) between two indicators were significantly correlated overall [R2 = 0.90, regional perfusion%, p < 0.001; R2 = 0.93, regional V/Q matching, p < 0.001] in 80 pairs of measurements (Fig. 2A and C). In the regional V/Q match, the LR and UL regions had a higher correlation of two indicators than the UR and LL regions. In perfusion distribution, LR region had the highest correlation. The Bland–Altman analysis showed a strong overall agreement in regional perfusion distribution(%) [mean bias 0.09% (95% confidence interval, CI: -0.91%, 1.09%), lower limits of agreement (LOA) -8.03% (95% CI: -9.75%, -6.31%), upper LOA 8.21% (95% CI: 6.48%, 9.93%)] and V/Q match (%) [mean bias − 0.04% (95% CI: -0.71%, 0.64%), lower LOA − 5.98% (95% CI: -7.14%, -4.82%), upper LOA 5.90% (95% CI: 4.75%, 7.06%)] between contrast agents in 160 pairs of measurements (Fig. 2B and D).

Fig. 2
figure 2

Correlation and Bland-Altman agreement analysis of four cross-regional perfusion distributions and V/Q match between hypertonic sodium saline and sodium bicarbonate. 2A, Scatter plot of regional lung perfusion (% of total perfusion) between indicators. Correlation defined by R2 is shown overall and within each lung region. 2B, Bland–Altman plot of regional lung perfusion (% of total perfusion) by lung region between indicators. 2C, Scatter plot of regional V/Q match (%) between indicators. 2D, Bland–Altman plot of r regional V/Q match (%) between indicators. LOA, limits of agreement

A case of repeated measurements

One patient received five pairs of contrast examinations on different days due to changes in his condition and the clinical need for lung perfusion assessment. A set of line graphs displaying changes in regional lung perfusion % and regional V/Q match % at different time points is presented in Fig. 3. In some regions, such as the lower right and upper right, the results from the two contrast agents are comparable. Whereas in certain regions, for example, the upper left and lower left, the regional lung perfusion % between NaHCO3 and NaCl are more differentiated.

Fig. 3
figure 3

Changes in regional lung perfusion % and regional V/Q match % by NaHCO3 and NaCl on different days in a single patient. NaHCO3, sodium bicarbonate; NaCl, sodium chloride

Discussion

To the best of our knowledge, this is the first clinical study to validate an alternative contrast agent for EIT-based regional lung perfusion assessment in clinical settings. In the present study, regional lung perfusion and the corresponding V/Q matching images were calculated using EIT with NaHCO3 and with NaCl in 16 patients with respiratory failure. Although the decrease in impedance introduced by NaHCO3 was lower than that by NaCl, the regional information provided by the two contrast agents was comparable. In a single patient who has received multiple examinations on different days, the EIT results generated by NaHCO3 and NaCl are similar in some lung areas and different in others. More evidence with larger samples is required to validate the agreement between the two agents in monitoring dynamic changes in patients’ pulmonary conditions in the future.

The EIT-based regional lung perfusion method is an emerging bedside tool for critically ill patients [17, 18]. It has been found to be a promising alternative to current lung perfusion imaging techniques such as single-photon emission computed tomography (SPECT), computed tomography (CT), or positron emission tomography (PET) [3, 4, 19]. The latter are often not applicable to ICU patients due to the risk of patient transportation. On the other hand, EIT enables timely and dynamic lung perfusion monitoring at the bedside, which provides valuable insights for clinicians. To date, hypertonic NaCl is the most frequently used contrast agent. The concentration of NaCl used by the researcher ranged from 3 to 20% (20% only presented in animal studies) [2]. Overall, 5-10% NaCl has been employed to assess lung perfusion and regional V/Q match in different clinical conditions at the bedside, such as ARDS, pulmonary embolism, COVID-19, etc [6, 16, 20,21,22,23,24]. The potential clinical applications of lung perfusion by EIT and advantages/disadvantages of 10% NaCl and 5% NaHCO3 were summarized in Table 2.

Table 2 Potential applications of EIT lung perfusion and advantages/disadvantages of 10% NaCl and 5% NaHCO3

The amount of fluid and chlorine contained in a bolus of hypertonic NaCl is small when compared to the whole internal environment. In general, the injected hypertonic NaCl is able to be equilibrated by the organism in a short period of time. However, in patients with severe illness in several conditions, the use of hypertonic NaCl is restricted. For example, in critically ill patients with hyperchloremia, which commonly occurs in septic shock, the addition of sodium chloride might deteriorate the situation and increase the risk of acute kidney injury [25]. Similarly, it is also prohibited in patients with severe electrolyte disorders, edematous diseases, and severe hypertension. Moreover, other side effects of hypertonic saline, such as global and regional hemodynamic effects [26], chloride overload, and impaired renal perfusion need to be considered. Therefore, it is warranted to validate another contrast agent without chloride in the critically ill patient with hyperchloremia.

NaHCO3 is commonly used and accessible in hospitals and ICUs. It is the basic treatment for severe metabolic acidemia with the ability to balance the pH of the individual organism [27]. Recently, NaHCO3 has been proposed as an alternative to hypertonic NaCl in two animal studies [12, 13]. Muders et al. investigated the ability of five different contrast agents in anesthetized pigs, including 5.85% NaCl, 8.4% NaHCO3, 5% Glucose, isotonic balanced crystalloid solution, and Iomeprol 400 mg/ml. Among them, NaCl and NaHCO3 showed the most satisfying results with the best signal and image quality. Gaulton et al. compared the proportion of lung perfusion distribution in different lung regions by 12% NaCl and 8.4% NaHCO3 in lung injury pig models. The results obtained with the two contrast agents were comparable. Thus, we conducted the first clinical research to further validate the potential of NaHCO3 as an alternative contrast agent in the EIT-based lung perfusion technique under clinical conditions. In the present study, we adopted 15 mL of 5% NaHCO3 rather than 10 mL of 8.4% NaHCO3 (used in previous animal studies) because it is more accessible in the hospital and potentially less irritating to patients. The two solutions are comparable in conductivity for they contain similar amounts of sodium ions. However, the selection of doses and concentrations requires further investigation.

As a contrast agent, NaHCO3 has several potential advantages: (1) NaHCO3 has no chloride ions. The associated risk of kidney injury might, therefore, be low. It has the advantage in patients with hyperchloremia. (2) NaHCO3 might have a lower impact on the internal environment, and multiple repeated bolus injections within a short period are relatively safe, especially in critically ill patients with shock. These patients have metabolic toxicity, and the administration of NaHCO3 is required. (3) NaHCO3 has a lower osmolality. It might cause less vascular irritation when performing peripheral bolus. Thus, NaHCO3 has the potential for peripheral venous injection, which might be able to eliminate the need for a central venous catheter.

There are several limitations in this study: (1) The sample size is small. (2) The change of plasmatic chloride, sodium levels, and pH were unavailable during different bolus solutions injections. The potential adverse effects (hyperchloremia, acidosis, and possible acute kidney injury) of a 10 ml bolus of 10% NaCl were speculated, and clear evidence was lacking. The included population was not sufficient to discern any differences following the injection of a single bolus per day. Hence, the potential benefits of 5% NaHCO3 (vs. 10% NaCl) require further study. (3) Compared to NaCl, the impact of the novel contrast agent (NaHCO3) on clinical decisions and outcomes was not evaluated. (4) We analyzed the agreement between the two contrast agents by dividing the lung into four regions, for it is most commonly used in clinical. A pixel-based analysis might be more accurate. (5) We did not use the cardiac output (CO) and minute ventilation (MV) for correction, which can also affect the analysis of V/Q. (6) End-expiratory and end-inspiratory holds were used during the saline bolus implementation for lung perfusion measurement. The end-expiratory hold was used in the present study, and the impaired circulation by breath hold was not found. The potential benefits of end-expiratory hold were the following: (a) little impact on venous return and circulation; (b) an obvious decrease of impedance could be easily caused at end-expiration with a lower impedance baseline. (7) The repeated measures were not preset in an individual patient. Actually, several lung perfusion assessments were required by the attending doctor on different days based on clinical conditions. We think an individual case with repeat measures might have interest in the dynamic assessment of agreement of the two solutions. (8) The measurement of lung perfusion by EIT is influenced not only by the contrast but also by the quality of injection (speed and consistency). For the saline bolus quality control, the investigator received the same standardized bolus training for the two contrast boluses. With the aim to reduce the variation of performers, the contrast bolus was done by the same investigator.

Conclusion

Our findings demonstrate that NaHCO3 may be a comparable indicator to measure regional lung perfusion and V/Q by EIT in a cohort of critically ill patients. Further study is required to validate the impact of NaHCO3 on clinical decisions and outcomes in clinical practice.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request. The datasets supporting the conclusions of this article are included within the article.

References

  1. He H, Chi Y, Long Y, Yuan S, Zhang R, Frerichs I, Möller K, Fu F, Zhao Z. Bedside evaluation of pulmonary embolism by saline contrast electrical impedance tomography method: A prospective observational study. Am J Respir Crit Care Med. 2020;202(10):1464–8.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Xu M, He H, Long Y. Lung perfusion assessment by bedside electrical impedance tomography in critically ill patients. Front Physiol. 2021;12:748724.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Borges JB, Suarez-Sipmann F, Bohm SH, Tusman G, Melo A, Maripuu E, Sandström M, Park M, Costa EL, Hedenstierna G, et al. Regional lung perfusion estimated by electrical impedance tomography in a piglet model of lung collapse. J Appl Physiol (1985). 2012;112(1):225–36.

    Article  PubMed  Google Scholar 

  4. Bluth T, Kiss T, Kircher M, Braune A, Bozsak C, Huhle R, Scharffenberg M, Herzog M, Roegner J, Herzog P, et al. Measurement of relative lung perfusion with electrical impedance and positron emission tomography: an experimental comparative study in pigs. Br J Anaesth. 2019;123(2):246–54.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Hentze B, Muders T, Luepschen H, Maripuu E, Hedenstierna G, Putensen C, Walter M, Leonhardt S. Regional lung ventilation and perfusion by electrical impedance tomography compared to single-photon emission computed tomography. Physiol Meas. 2018;39(6):065004.

    Article  PubMed  Google Scholar 

  6. Spinelli E, Kircher M, Stender B, Ottaviani I, Basile MC, Marongiu I, Colussi G, Grasselli G, Pesenti A, Mauri T. Unmatched ventilation and perfusion measured by electrical impedance tomography predicts the outcome of ARDS. Crit Care. 2021;25(1):192.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Slobod D, Damia A, Leali M, Spinelli E, Mauri T. Pathophysiology and clinical meaning of Ventilation-Perfusion mismatch in the acute respiratory distress syndrome. Biology (Basel) 2022, 12(1).

  8. Karsten J, Grusnick C, Paarmann H, Heringlake M, Heinze H. Positive end-expiratory pressure Titration at bedside using electrical impedance tomography in post-operative cardiac surgery patients. Acta Anaesthesiol Scand. 2015;59(6):723–32.

    Article  CAS  PubMed  Google Scholar 

  9. Bito K, Shono A, Kimura S, Maruta K, Omoto T, Aoki A, Oe K, Kotani T. Clinical implications of determining individualized positive End-Expiratory pressure using electrical impedance tomography in Post-Cardiac surgery patients: A prospective, Non-Randomized interventional study. J Clin Med 2022, 11(11).

  10. Semler MW, Self WH, Wanderer JP, Ehrenfeld JM, Wang L, Byrne DW, Stollings JL, Kumar AB, Hughes CG, Hernandez A, et al. Balanced crystalloids versus saline in critically ill adults. N Engl J Med. 2018;378(9):829–39.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Zampieri FG, Cavalcanti AB, Di Tanna GL, Damiani LP, Hammond NE, Machado FR, Micallef S, Myburgh J, Ramanan M, Venkatesh B, et al. Balanced crystalloids versus saline for critically ill patients (BEST-Living): a systematic review and individual patient data meta-analysis. Lancet Respir Med. 2024;12(3):237–46.

    Article  PubMed  Google Scholar 

  12. Muders T, Hentze B, Leonhardt S, Putensen C. Evaluation of different contrast agents for regional lung perfusion measurement using electrical impedance tomography: an experimental pilot study. J Clin Med 2023, 12(8).

  13. Gaulton TG, Martin K, Xin Y, Victor M, Ribeiro De Santis Santiago R, Britto Passos Amato M, Berra L, Cereda M. Regional lung perfusion using different indicators in electrical impedance tomography. J Appl Physiol (1985). 2023;135(3):500–7.

    Article  CAS  PubMed  Google Scholar 

  14. Zhao Z, Yun PJ, Kuo YL, Fu F, Dai M, Frerichs I, Möller K. Comparison of different functional EIT approaches to quantify tidal ventilation distribution. Physiol Meas. 2018;39(1):01nt01.

    Article  PubMed  Google Scholar 

  15. He H, Long Y, Frerichs I, Zhao Z. Detection of acute pulmonary embolism by electrical impedance tomography and saline bolus injection. Am J Respir Crit Care Med. 2020;202(6):881–2.

    Article  PubMed  Google Scholar 

  16. He H, Chi Y, Long Y, Yuan S, Zhang R, Yang Y, Frerichs I, Möller K, Fu F, Zhao Z. Three broad classifications of acute respiratory failure etiologies based on regional ventilation and perfusion by electrical impedance tomography: a hypothesis-generating study. Ann Intensive Care. 2021;11(1):134.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  17. Frerichs I, Amato MB, van Kaam AH, Tingay DG, Zhao Z, Grychtol B, Bodenstein M, Gagnon H, Böhm SH, Teschner E, et al. Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the translational EIT development study group. Thorax. 2017;72(1):83–93.

    Article  PubMed  Google Scholar 

  18. Smit HJ, Vonk Noordegraaf A, Marcus JT, Boonstra A, de Vries PM, Postmus PE. Determinants of pulmonary perfusion measured by electrical impedance tomography. Eur J Appl Physiol. 2004;92(1–2):45–9.

    Article  PubMed  Google Scholar 

  19. Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Quintel M, Hellige G. Regional lung perfusion as determined by electrical impedance tomography in comparison with electron beam CT imaging. IEEE Trans Med Imaging. 2002;21(6):646–52.

    Article  PubMed  Google Scholar 

  20. Grassi LG, Santiago R, Florio G, Berra L. Bedside evaluation of pulmonary embolism by electrical impedance tomography. Anesthesiology. 2020;132(4):896.

    Article  PubMed  Google Scholar 

  21. Prins SA, Weller D, Labout JAM, den Uil CA. Electrical impedance tomography as a bedside diagnostic tool for pulmonary embolism. Crit Care Explor. 2023;5(1):e0843.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Safaee Fakhr B, Araujo Morais CC, De Santis Santiago RR, Di Fenza R, Gibson LE, Restrepo PA, Chang MG, Bittner EA, Pinciroli R, Fintelmann FJ, et al. Bedside monitoring of lung perfusion by electrical impedance tomography in the time of COVID-19. Br J Anaesth. 2020;125(5):e434–6.

    Article  CAS  PubMed  Google Scholar 

  23. Mauri T, Spinelli E, Scotti E, Colussi G, Basile MC, Crotti S, Tubiolo D, Tagliabue P, Zanella A, Grasselli G, et al. Potential for lung recruitment and Ventilation-Perfusion mismatch in patients with the acute respiratory distress syndrome from coronavirus disease 2019. Crit Care Med. 2020;48(8):1129–34.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. Perier F, Tuffet S, Maraffi T, Alcala G, Victor M, Haudebourg AF, De Prost N, Amato M, Carteaux G, Mekontso Dessap A. Effect of positive End-Expiratory pressure and proning on ventilation and perfusion in COVID-19 acute respiratory distress syndrome. Am J Respir Crit Care Med. 2020;202(12):1713–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  25. Suetrong B, Pisitsak C, Boyd JH, Russell JA, Walley KR. Hyperchloremia and moderate increase in serum chloride are associated with acute kidney injury in severe sepsis and septic shock patients. Crit Care. 2016;20(1):315.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Kien ND, Kramer GC. Cardiac performance following hypertonic saline. Braz J Med Biol Res. 1989;22(2):245–8.

    CAS  PubMed  Google Scholar 

  27. Fujii T, Udy A, Licari E, Romero L, Bellomo R. Sodium bicarbonate therapy for critically ill patients with metabolic acidosis: A scoping and a systematic review. J Crit Care. 2019;51:184–91.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was supported by the CAMS Innovation Fund for Medical Sciences(Grant No 2023-I2M-C&T-B-014)

, National Natural Science Foundation of China (Grant No 82272249), the National Key Research and Development Program (2022YFC2404805).

Author information

Authors and Affiliations

Authors

Contributions

HH and YC conceived the study protocol; HH, YC, YL, SY, and YG participated in the design and coordination of the study; YH, HH, YC, YG, SY, SW, and YL collected study data; YG, YC, YH, HH, YL, SY, SW, and ZZ participated in data interpretation; YG, YH, HH and YC drafted the manuscript; YL, SY and ZZ revised the manuscript significantly. All authors read and approved the final version of the manuscript.

Corresponding author

Correspondence to Huaiwu He.

Ethics declarations

Ethics approval and consent to participate

This study was conducted in accordance with the principles outlined in the Declaration of Helsinki. The study protocol was approved by the Institutional Research and Ethics Committee of the Peking Union Medical College Hospital (I-24PJ0505) and is registered at ClinicalTrials.gov (NCT06868810). Written informed consent was obtained from all participants or their next of kin.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gao, Y., He, Y., Chi, Y. et al. Comparison of 5% sodium bicarbonate and 10% sodium chloride as contrast agents for lung perfusion with electrical impedance tomography: a prospective clinical study. BMC Pulm Med 25, 190 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12890-025-03665-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12890-025-03665-2

Keywords